Reliability of respiratory-triggered two-dimensional cine k-adaptive-t-autocalibrating reconstruction for Cartesian sampling for the assessment of biventricular volume and function in patients with repaired tetralogy of Fallot
- Source :
- The British Journal of Radiology
- Publication Year :
- 2021
-
Abstract
- Objective: To compare left ventricular (LV) and right ventricular (RV) volume, function, and image quality of a respiratory-triggered two-dimensional (2D)-cine k-adaptive-t-autocalibrating reconstruction for Cartesian sampling (2D kat-ARC) with those of the standard reference, namely, breath-hold 2D balanced steady-state free precession (2D SSFP), in patients with repaired tetralogy of Fallot (TOF). Methods: 30 patients (14 males, mean age 32.2 ± 13.9 years) underwent cardiac magnetic resonance, and 2D kat-ARC and 2D SSFP images were acquired on short-axis view. Biventricular end-diastolic volume (EDV) and end-systolic volume (ESV), stroke volume (SV), ejection fraction (EF), and LV mass (LVM) were analysed. Results: The 2D kat-ARC had significantly shorter scan time (35.2 ± 9.1 s vs 80.4 ± 16.7 s; p < 0.0001). Despite an analysis of image quality showed significant impairment using 2D kat-ARC compared to 2D SSFP cine (p < 0.0001), the two sequences demonstrated no significant difference in terms of biventricular EDV, LVESV, LVSV, LVEF, and LVM. However, the RVESV was overestimated for 2D kat-ARC compared with that for 2D SSFP (73.8 ± 43.2 ml vs 70.3 ± 44.5 ml, p = 0.0002) and the RVSV and RVEF were underestimated (RVSV = 46.2±20.5 ml vs 49.4 ± 20.4 ml, p = 0.0024; RVEF = 40.2±12.7% vs. 43.5±14.0%, p = 0.0002). Conclusion: Respiratory-triggered 2D kat-ARC cine is a reliable technique that could be used in the evaluation of LV volumes and function. Advances in knowledge: 2D cine kat-ARC is a reliable technique for the assessment LV volume and function in patients with repaired TOF.
- Subjects :
- Adult
Male
medicine.medical_specialty
Image quality
Heart Ventricles
Ventricular Dysfunction, Right
030204 cardiovascular system & hematology
030218 nuclear medicine & medical imaging
law.invention
Breath Holding
03 medical and health sciences
Ventricular Dysfunction, Left
0302 clinical medicine
Postoperative Complications
Sampling (signal processing)
law
Internal medicine
Image Interpretation, Computer-Assisted
medicine
Humans
Radiology, Nuclear Medicine and imaging
Cartesian coordinate system
Respiratory system
Tetralogy of Fallot
Full Paper
business.industry
Reproducibility of Results
Stroke Volume
General Medicine
Function (mathematics)
Stroke volume
medicine.disease
Magnetic Resonance Imaging
Cardiology
Female
business
Volume (compression)
Subjects
Details
- ISSN :
- 1748880X
- Volume :
- 94
- Issue :
- 1120
- Database :
- OpenAIRE
- Journal :
- The British journal of radiology
- Accession number :
- edsair.doi.dedup.....e70c666e44db9f4019b51d3b20128540