Back to Search Start Over

Quantitative read-out of Al2O3:C,Mg-based fluorescent nuclear track detectors using a commercial confocal microscope

Authors :
Greilich, Steffen
Osinga, Julia-Maria
Niklas, Martin
Lauer, Florian
Bestvater, Felix
Jäkel, Oliver
Publication Year :
2014
Publisher :
arXiv, 2014.

Abstract

Fluorescent nuclear track detectors (FNTD) show great potential for applications in ion-beam therapy research, such as dosimetry, advanced beam characterization, in-vivo use or as radiobiological assay. A essential feature of FNTDs is their ability to assess the energy loss of single ions yielding for example LET estimations. This article describes the basic characterisations of FNTDs and our read-out system (a Zeiss LSM710 confocal laser scanning microscope) to enable quantative measurements of energy loss.<br />Comment: 13 pages, 10 figures

Details

Database :
OpenAIRE
Accession number :
edsair.doi.dedup.....e9973eef1f1aba877933e931c51a7fbf
Full Text :
https://doi.org/10.48550/arxiv.1407.0986