Failure of the Durom MetasulĀ® Acetabular Component

Authors :
William T. Long
Manish Dastane
Michael J. Harris
Lawrence D. Dorr
Zhinian Wan
Publication Year :
2009
Publisher :
Springer-Verlag, 2009.

Abstract

Large-diameter metal-on-metal articulations reportedly provide better stability and range of motion than smaller diameter bearings. We therefore asked whether a large-diameter (44- to 50-mm) metal-on-metal articulation (Durom) would eliminate dislocation and provide similar functional scores and clinical and radiographic failure rates as those with 28-mm articulation. We prospectively followed 181 patients (207 hips) who had a large-diameter articulation implanted between May 2006 and November 2007. We compared these patients with a historical control of 54 patients who had a small-diameter (28-mm Metasul) articulation. All patients had a Harris hip score and a self-assessment of outcome and radiographic followup. The minimum followup was 1 year (mean, 1.6 years; range, 1-2 years). During the followup period, we performed revisions on 29 patients (30 hips [15%]) with 21 of 29 (72%) having radiographic criteria of loosening. Thirteen retrieved cups and acetabular tissue were examined histologically. Twenty-eight of 151 unrevised patients had radiographic impending failure; 12 without revision had clinical failure. Eight patients (nine hips) had both clinical failure and impending radiographic failure. Cup inclination was 41.3 degrees +/- 5.4 degrees and anteversion was 20.2 degrees +/- 7 degrees. The revision rate and quality of clinical results were unacceptable as compared with our historical controls. We do not recommend use of the Durom implant.Level IV, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence.

Details

Language :
English
Database :
OpenAIRE
Accession number :
edsair.doi.dedup.....ec7dc8d433a9cb0ed5fe839878ec9be7