Back to Search
Start Over
Diffusion-Weighted Echo Planar Imaging Using Multiplexed Sensitivity Encoding and Reverse Polarity Gradient in Head Andneck Cancer: An Initial Study
- Source :
- Tomography, Volume 6, Issue 2, Pages 231-240, Tomography; Volume 6; Issue 2; Pages: 231-240
- Publication Year :
- 2020
- Publisher :
- MDPI AG, 2020.
-
Abstract
- We aimed to compare the geometric distortion (GD) correction performance and apparent diffusion coefficient (ADC) measurements of single-shot diffusion-weighted echo-planar imaging (SS-DWEPI), multiplexed sensitivity encoding (MUSE)-DWEPI, and MUSE-DWEPI with reverse-polarity gradient (RPG) in phantoms and patients. We performed phantom studies at 3T magnetic resonance imaging (MRI) using the American College of Radiology phantom and Quantitative Imaging Biomarker Alliance DW-MRI ice-water phantom to assess GD and effect of distortion in the measurement of ADC, respectively. Institutional review board approved the prospective clinical component of this study. DW-MRI data were obtained from 11 patients with head and neck cancer using these three DW-MRI methods. Wilcoxon signed-rank (WSR) and Kruskal–Wallis (KW) tests were used to compare ADC values, and qualitative rating by radiologist between three DW-MRI methods. In the ACR phantom, GD of 0.17% was observed for the b = 0 s/mm2 image of the MUSE-DWEPI with RPG method compared with that of 1.53% and 2.1% of MUSE-DWEPI and SS-DWEPI, respectively<br />The corresponding methods root-mean-square errors were 0.58, 3.37, and 5.07 mm. WSR and KW tests showed no significant difference in the ADC measurement between these three DW-MRI methods for both healthy masseter muscles and neoplasms (P &gt<br />05). We observed improvement in spatial accuracy for MUSE-DWEPI with RPG in the head and neck region with a higher correlation (R2 = 0.791) compared with that for SS-DWEPI (R2 = 0.707) and MUSE-DWEPI (R2 = 0.745). MUSE-DWEPI with RPG significantly reduces the distortion compared with MUSE-DWEPI or conventional SS-DWEPI techniques, and the ADC values were similar.
- Subjects :
- Materials science
magnetic field inhomogeneity
Wilcoxon signed-rank test
eddy current
Imaging phantom
head and neck region
diffusion-weighted MRI
echo planar imaging
geometrical distortion
multiplexed sensitivity encoding
reverse-polarity gradients
Distortion
medicine
Humans
Effective diffusion coefficient
Radiology, Nuclear Medicine and imaging
Prospective Studies
Diffusion (business)
Research Articles
medicine.diagnostic_test
Echo-Planar Imaging
Phantoms, Imaging
business.industry
Head and neck cancer
Magnetic resonance imaging
medicine.disease
body regions
Diffusion Magnetic Resonance Imaging
Head and Neck Neoplasms
Nuclear medicine
business
Diffusion MRI
Subjects
Details
- ISSN :
- 2379139X
- Volume :
- 6
- Database :
- OpenAIRE
- Journal :
- Tomography
- Accession number :