Back to Search Start Over

Implementation and assessment of the black body bias correction in quantitative neutron imaging

Authors :
Muriel Siegwart
Christian Gruenzweig
M. Morgano
Eberhard Lehmann
Anders Kaestner
Jan Hovind
Markus Strobl
Pierre Boillat
Chiara Carminati
Peter Vontobel
David Mannes
Florian Schmid
Marc Raventós
Pavel Trtik
Source :
PLOS ONE, PLoS ONE, Vol 14, Iss 1, p e0210300 (2019), PLoS ONE

Abstract

We describe in this paper the experimental procedure, the data treatment and the quantification of the black body correction: an experimental approach to compensate for scattering and systematic biases in quantitative neutron imaging based on experimental data. The correction algorithm is based on two steps; estimation of the scattering component and correction using an enhanced normalization formula. The method incorporates correction terms into the image normalization procedure, which usually only includes open beam and dark current images (open beam correction). Our aim is to show its efficiency and reproducibility: we detail the data treatment procedures and quantitatively investigate the effect of the correction. Its implementation is included within the open source CT reconstruction software MuhRec. The performance of the proposed algorithm is demonstrated using simulated and experimental CT datasets acquired at the ICON and NEUTRA beamlines at the Paul Scherrer Institut.

Details

Language :
English
ISSN :
19326203
Volume :
14
Issue :
1
Database :
OpenAIRE
Journal :
PLOS ONE
Accession number :
edsair.doi.dedup.....f0e5f13c77fa33227cc2da4b8ceda1f9
Full Text :
https://doi.org/10.1371/journal.pone.0210300