Back to Search Start Over

LASIK excimer misfire on fluence test mode: induced refractive error quantification and corneal ablation profile

Authors :
Priyanka Prasad
Avilasha Mohapatra
Sudarshan Khokhar
Amber Amar Bhayana
Source :
BMJ case reports. 14(7)
Publication Year :
2023

Abstract

A 23-year-old man with refractive error as in [figure 1A][1] was planned for LASIK (Laser assisted in-situ keratomileusis) refractive surgery (ZEISS MEL 90). After flap preparation, the subject was accidentally fired on by excimer laser at fluence test settings[1][2] instead of that for the

Details

ISSN :
1757790X
Volume :
14
Issue :
7
Database :
OpenAIRE
Journal :
BMJ case reports
Accession number :
edsair.doi.dedup.....f1cde7f259a22e9092c7b9b2a56ac465