Evaluation of venous plasma glucose measured by point-of-care testing (Accu-Chek Inform II) and a hospital laboratory hexokinase method (Cobas c701) in oral glucose tolerance testing during pregnancy - a challenge in diagnostic accuracy

Authors :
Eva Landberg
Anna Norling
Sofia Nevander
Mohammed Hadi
Marie Blomberg
Bertil Ekman
Caroline Lilliecreutz
Source :
Scandinavian journal of clinical and laboratory investigation. 81(8)
Publication Year :
2021

Abstract

To diagnose gestational diabetes mellitus (GDM), plasma glucose measurements during oral glucose tolerance test (OGTT) put high demands on the methods in terms of accuracy. The aim was to evaluate and compare diagnostic performance of a point-of-care test and a glucose hexokinase laboratory method. Using risk-based screening, 175 pregnant women were included. They underwent a 75 g OGTT in their 28th (median) week of gestation. Venous blood was collected in two different tubes. Plasma glucose was measured on Cobas c701 and in duplicates on AccuChek Inform II (both methods from Roche Diagnostics). Accuracy was assessed by participating in external control programs with reference method assigned values. The methods were compared for all samples (