Back to Search Start Over

Chiral analysis of methadone in plasma by high-performance liquid chromatography

Authors :
Olof Beck
Pierre Lafolie
Lars O. Boreus
Gun Jacobsson
Source :
Journal of chromatography. 570(1)
Publication Year :
1991

Abstract

A method for the chiral high-performance liquid chromatographic analysis of methadone in plasma has been developed. The method employed organic solvent extraction, enantiomeric separation on a Chiral AGP column, and ultraviolet absorption detection at 212 nm. The intra-day variation in the quantification of methadone enantiomers was less than 9% at the 100 ng/ml level, and the values obtained correlated well with those from a gas chromatographic-mass spectrometric method. Results from patients indicate inter- and intra-individual differences in the ratio between l- and d-methadone in plasma during therapy with racemic methadone. In one patient, a higher level of d-methadone in plasma was caused by both faster elimination and lower bioavailability of l-methadone.

Details

Volume :
570
Issue :
1
Database :
OpenAIRE
Journal :
Journal of chromatography
Accession number :