Abstract
- This paper describes the beam characteristics and dosimetry measurements performed on the 6-MV photon beam of a new model of linear accelerator, three of which were recently introduced and installed in our institution. Percent depth dose and tissue maximum ratio tables for a variety of field sizes and depths, as well as other parameters used for treatment planning are presented. These accelerators are the first of their kind using both hardware and software tools to control interlocks. Checking procedures for these interlocks are available from the authors upon request. Comparison of characteristic parameters between these three new 6-MV linear accelerators and with the 6-MV beams of two other accelerators is also made.