Ablation zone geometry after CT-guided hepatic microwave ablation: Evaluation of a semi-automatic software and comparison of two different ablation systems

Authors :
Vo Chieu, V.D.
Wacker, F.
Rieder, C.
Pöhler, G.H.
Schumann, C.
Ballhausen, H.
Ringe, K.I.
Publica
Publication Year :
2020

Abstract

Purpose: The aims of this study were to evaluate a semi-automatic segmentation software for assessment of ablation zone geometry in computed tomography (CT)-guided microwave ablation (MWA) of liver tumors and to compare two different MWA systems. Material and Methods: 27 patients with 40 hepatic tumors (primary liver tumor n = 20, metastases n = 20) referred for CT-guided MWA were included in this retrospective IRB-approved study. MWA was performed using two systems (system 1: 915 MHz; n = 20; system 2: 2.45 GHz; n = 20). Ablation zone segmentation and ellipticity index calculations were performed using SAFIR (Software Assistant for Interventional Radiology). To validate semi-automatic software calculations, results (2 perpendicular diameters, ellipticity index, volume) were compared with those of manual analysis (intraclass correlation, Pearson's correlation, Mann-Whitney U test; p