Back to Search
Start Over
Development and test of the high-precision gamma-beam monitor at the ESFRI facility ELI-NP
- Publication Year :
- 2017
- Publisher :
- Publications Office of the European Union, 2017.
-
Abstract
- This document describes the development of a fission chamber for gamma beam intensity monitoring at the high-energy gamma line at ELI-NP. The fission chamber foreseen for monitoring the gamma beam intensity on the high-energy beam line at ELI-NP is an ionization chamber with several 238U deposits of a thickness in the range from 100 to 150 μg/cm2 in a compact design. The innovative design of the chamber allows measuring for each individual micro pulse arriving at a time interval of 1.6x10^-8 s.<br />JRC.G.2-Standards for Nuclear Safety, Security and Safeguards
Details
- Language :
- English
- Database :
- OpenAIRE
- Accession number :
- edsair.od......1739..ffa2c5b81d8c72c03aae2e9a51ff4e4a