Back to Search Start Over

A digitizer with exceptional accuracy for use in prosthetics research: a technical note

Authors :
Joan E, Sanders
Stuart B, Mitchell
Santosh G, Zachariah
Kelvin, Wu
Source :
Journal of rehabilitation research and development. 40(2)
Publication Year :
2004

Abstract

A mechanical digitizer was developed for use in prosthetics research where measurements of small differences in shape are of interest. Root-mean-square error was 0.075 mm in the radial direction, 0.05 degrees in the tangential direction, and 0.1 mm in the vertical direction. The system has potential use for time-dependent assessment of changes in socket and residuum cast shape, assessment of socket fabrication systems, and development of accurate prosthetic finite element models.