Acute Gain in Minimal Lumen Area Following Implantation of Everolimus-Eluting ABSORB Biodegradable Vascular Scaffolds or Xience Metallic Stents: Intravascular Ultrasound Assessment From the ABSORB II Trial

Authors :
Yohei, Sotomi
Yuki, Ishibashi
Pannipa, Suwannasom
Shimpei, Nakatani
Yun-Kyeong, Cho
Maik J, Grundeken
Yaping, Zeng
Hiroki, Tateishi
Pieter C, Smits
Paul, Barragan
Ran, Kornowski
Anthony H, Gershlick
Stephan, Windecker
Robert-Jan, van Geuns
Antonio L, Bartorelli
Robbert J, de Winter
Jan, Tijssen
Patrick W, Serruys
Yoshinobu, Onuma
Source :
JACC. Cardiovascular interventions. 9(12)
Publication Year :
2015

Abstract

The study compared, by intravascular ultrasound (IVUS), acute gain (AG) at the site of the pre-procedural minimal lumen area (MLA) achieved by either the Absorb (Abbott Vascular, Santa Clara, California) scaffold or the Xience stent and identified the factors contributing to the acute performance of these devices.It is warranted that the acute performance of Absorb matches that of metallic stents; however, concern exists about acute expansion and lumen gain with the use of Absorb.Of a total of 501 patients (546 lesions) in the ABSORB II (ABSORB II Randomized Controlled Trial) randomized trial, 445 patients with 480 lesions were investigated by IVUS pre- and post-procedure. Comparison of MLA pre- and post-procedure was performed at the MLA site by matching pre- and post-procedural IVUS pullbacks.Lower AG on IVUS (lowest tertile) occurred more frequently in the Absorb arm than in the Xience arm (3.46 mm(2) vs. 4.27 mm(2), respectively; p 0.001; risk ratio: 3.04; 95% confidence interval: 1.94 to 4.76). The plaque morphology at the MLA cross-section was not independently associated with IVUS acute gain. The main difference in AG in MLD by angiography was observed at the time of device implantation (Xience vs. Absorb, Δ+1.50 mm vs. Δ+1.23 mm, respectively), whereas the gain from post-dilation was similar between the 2 arms (Δ+0.16 mm vs. Δ+0.16 mm) when patients underwent post-dilation, although expected balloon diameter was smaller in the Absorb arm than in the Xience arm (p = 0.003) during post-dilation.At the site of the pre-procedural MLA, the increase of the lumen post-procedure was smaller in the Absorb-arm than in the Xience arm. To achieve equivalent AG to Xience, the implantation of Absorb may require more aggressive strategies at implantation, pre- and post-dilation than the technique used in the ABSORB II trial. (ABSORB II Randomized Controlled Trial [ABSORB II]; NCT01425281).

Details

ISSN :
18767605
Volume :
9
Issue :
12
Database :
OpenAIRE
Journal :
JACC. Cardiovascular interventions
Accession number :
edsair.pmid..........10a2e0dedd4f77f93eb4f99b5ed7ecec