Kinetic determination of serum haptoblobin with a centrifugal analyzer
- Source :
- Clinical chemistry. 22(12)
- Publication Year :
- 1976
-
Abstract
- We describe a method for determining haptoglobin with a centrifugal analyzer that is based on haptoglobin combining stoichiometrically with hemoglobin to form a complex that has peroxidase-like activity proportional to the quantity of haptoglobin present. Under assay conditions, unbound hemoglobin exhibits only a small fraction of the total peroxidase activity. Activity is measured colorimetrically at 405 nm after reaction with o-dianisdine and ethyl hydrogen peroxide. The procedure is standardized by saturating aknown amount of hemoglobin with a serum whose hemoglobin binding capacity exceeds the amount of hemoglobin in the assay system. The mean and mean within-run precision of our method, determined by performing 17 replicate assays of both a pooled normal serum and a 10-fold dilution of the serum, was 1.13 g/liter (CV, 2.9%), and 106 mg/liter (cv, 5.8%), respectively. The 95 percentile estimate of the normal range by our method is 0.45-1.85 g/liter hemoglobin binding capacity. When results by our automated method were compared to those by a manual method [Scand. J. Clin. Lab. 2nvest. 18, 80 (1965)], the slope of the unweighted linear least-squares regression line was .970 the y-intercept 26 mg/liter, and the correlation coefficient .995.
- Subjects :
- Kinetics
Autoanalysis
Haptoglobins
Peroxidases
Humans
Hemoglobin A
Hemolysis
Subjects
Details
- ISSN :
- 00099147
- Volume :
- 22
- Issue :
- 12
- Database :
- OpenAIRE
- Journal :
- Clinical chemistry
- Accession number :
- edsair.pmid..........a055f671ddb890620cb315d6cf35d862