Back to Search
Start Over
[Radiotherapy Dosimeter Calibration for Absorbed Dose to Water Using Medical Accelerators]
- Source :
- Igaku butsuri : Nihon Igaku Butsuri Gakkai kikanshi = Japanese journal of medical physics : an official journal of Japan Society of Medical Physics. 41(3)
- Publication Year :
- 2021
-
Abstract
- A calibration service using a medical accelerator has been launched to calibrate a radiotherapy dosimeter in terms of an absorbed dose to water. The radiotherapy dosimeter calibrated by the calibration service can measure the absorbed dose to water without a beam quality conversion factor. In this paper, an overview of the calibration service for a high-energy photon beam and a high-energy electron beam was described, as well as methods of absorbed dose measurement and cross-calibration using the calibrated radiotherapy dosimeter. And the development status of a dose standard for a particle beam was reported.
Details
- ISSN :
- 13455354
- Volume :
- 41
- Issue :
- 3
- Database :
- OpenAIRE
- Journal :
- Igaku butsuri : Nihon Igaku Butsuri Gakkai kikanshi = Japanese journal of medical physics : an official journal of Japan Society of Medical Physics
- Accession number :
- edsair.pmid..........b4dc6bc820fb3cc2851508d47a8282dc