Identifying the source of spurious signals caused by B
- Source :
- Magnetic resonance in medicine. 88(1)
- Publication Year :
- 2022
-
Abstract
- Single-voxel MRS (SV MRS) requires robust volume localization as well as optimized crusher and phase-cycling schemes to reduce artifacts arising from signal outside the volume of interest. However, due to local magnetic field gradients (BWe investigated all possible coherence pathways associated with imperfect localization in a semi-localized by adiabatic selective refocusing (semi-LASER) sequence for potential rephasing of signals arising from unwanted coherence pathways by a local magnetic field gradient. We searched for locations in the BOur analysis revealed that potential sources of spurious signals were scattered over multiple locations throughout the brain. This was confirmed by 3D-CSI data. Moreover, we showed that the number of potential locations where spurious signals could originate from monotonically decreases with crusher strength.We proposed a method to identify the source of spurious signals in SV
- Subjects :
- Magnetic Fields
Phantoms, Imaging
Brain
Humans
Artifacts
Magnetic Resonance Imaging