Back to Search Start Over

Effect of hematocrit on dialyzer urea and creatinine clearance indices in a hemodialysis patient with erythrocytosis

Authors :
A C, Nwosu
A H, Tzamaloukas
D, Malhotra
M C, Saddler
G H, Murata
Source :
The International journal of artificial organs. 17(11)
Publication Year :
1994

Abstract

The association between azotemic index dialyzer clearances and hematocrit was investigated in a 63-year-old dialysis-dependent man with acquired renal cysts. During 43 months of hemodialysis, hematocrit rose from 27.3 to 65.0 vol%, as a consequence of high serum erythropoietin levels. Concomitantly, dry weight also increased from 116.8 to 140.8 kg. Both hematocrit and dry weight correlated with: (a) urea reduction ratio, (b) creatinine reduction ratio (CRR), and (c) KT/V urea. All correlations were negative. Stepwise regression showed that only hematocrit was an independent correlate of the CRR (CRR = 0.662 - 007* Hct, R2 = 0.770); whereas, both hematocrit (Hct) and weight (W) were independent correlates of KT/V urea (KT/V = 2.070 - 0.005*Hct - 0.009*W,R2 = 0.721). In addition to creatinine clearance, urea clearance through the dialyzer is reduced by a rising hematocrit. The effect of hematocrit on urea clearance is relatively small. Therefore, it requires large changes in hematocrit in order to be detected.

Details

ISSN :
03913988
Volume :
17
Issue :
11
Database :
OpenAIRE
Journal :
The International journal of artificial organs
Accession number :
edsair.pmid..........fcd7133c6d2f1d784206100f532e65b4