Back to Search Start Over

Geiger Mode APD performance in a cryogenic two-phase Ar avalanche detector based on THGEMs

Authors :
Bondar, A.
Buzulutskov, A.
Grebenuk, A.
Sokolov, A.
Akimov, D.
Alexandrov, I.
Breskin, A.
Source :
Nucl.Instrum.Meth.A628:364-368,2011
Publication Year :
2010

Abstract

Characteristic properties of a Geiger Mode APD (G-APD) in a THGEM-based cryogenic two-phase Ar avalanche detector were studied in view of potential applications in rare-event experiments. G-APD signal amplitude and noise characteristics at cryogenic temperatures turned out to be superior to those at room temperature. The effective detection of avalanche scintillations from THGEM-multiplier holes in two-phase Ar has been demonstrated using a G-APD without wavelength shifter. At an avalanche gain of 60, the avalanche scintillation yield measured by the G-APD was as high as 0.9 photoelectrons per avalanche electron, extrapolated to 4pi acceptance.<br />Comment: 4 pages, 8 figures. Presented at Vienna Conference on Instrumentation (Feb 15-20, 2010, Vienna, Austria). Submitted to the Proceedings

Details

Database :
arXiv
Journal :
Nucl.Instrum.Meth.A628:364-368,2011
Publication Type :
Report
Accession number :
edsarx.1003.1597
Document Type :
Working Paper
Full Text :
https://doi.org/10.1016/j.nima.2010.07.002