Experimental study of THGEM detector with mini-rim

Authors :
Zhang, Ai-Wu
Yu, Bo-Xiang
Xie, Yu-Guang
Liu, Hong-Bang
An, Zheng-Hua
Wang, Zhi-Gang
Cai, Xiao
Sun, Xi-Lei
Shi, Feng
Fang, Jian
Xue, Zhen
Lu, Qi-Wen
Sun, Li-Jun
Ge, Yong-Shuai
Liu, Ying-Biao
Hu, Tao
Zhou, Li
Lu, Jun-Guang
Publication Year :
2011

Abstract

The gas gain and energy resolution of single and double THGEM detectors (5{\times}5cm2 effective area) with mini-rims (rim is less than 10{\mu}m) were studied. The maximum gain can reach 5{\times}103 and 2{\times}105 for single and double THGEM respectively, while the energy resolution of 5.9 keV X-ray varied from 18% to 28% for both single and double THGEM detectors of different hole sizes and thicknesses.All the experiments were investigated in mixture of noble gases(argon,neon) and small content of other gases(iso-butane,methane) at atmospheric pressure.<br />Comment: 4pages,6figures, it has been submitted to Chinese Physics C