Back to Search Start Over

A study of pickup and signal processing for HLS-II bunch current measurement system

Authors :
Yong-Liang, Yang
Tian-Ji, Ma
Bao-Gen, Sun
Ji-Gang, Wang
Jun-Ying, Zou
Chao-Cai, Cheng
Ping, Lu
Publication Year :
2013

Abstract

For the HLS-II bunch current measurement system, in order to obtain the absolute value of bunch current, the calibration factor should be determined by using DCCT. At the HLS storage ring, the stretch effect of bunch length is observed and the change rate is about 19% when the bunch current decays over time and this will affect the performance of bunch current detection. To overcome the bunch stretch influence in the HLS-II bunch current measurement, an evaluation about pickup type and signal processing is carried out. Strip-line pickup and button pickup are selectable, and the theoretical analysis and demonstration experiment are performed to find out an acceptable solution for the bunch current measurement system at HLS-II. The experimental data analysis shows that the normalized calibration factor will change by about 27% when the bunch length change by about 19% if using the button pickup and processing by peak value of bunch signal, the influence will be reduced to less 2% if adopting the strip-line pickup and integral.<br />Comment: 5 pages, 8 figures

Subjects

Subjects :
Physics - Accelerator Physics

Details

Database :
arXiv
Publication Type :
Report
Accession number :
edsarx.1304.4706
Document Type :
Working Paper