High-precision efficiency calibration of a high-purity co-axial germanium detector
- Publication Year :
- 2014
-
Abstract
- A high-purity co-axial germanium detector has been calibrated in efficiency to a precision of about 0.15% over a wide energy range. High-precision scans of the detector crystal and gamma-ray source measurements have been compared to Monte-Carlo simulations to adjust the dimensions of a detector model. For this purpose, standard calibration sources and short-lived on-line sources have been used. The resulting efficiency calibration reaches the precision needed e.g. for branching ratio measurements of super-allowed beta decays for tests of the weak-interaction standard model.
- Subjects :
- Physics - Instrumentation and Detectors
Nuclear Experiment
Subjects
Details
- Database :
- arXiv
- Publication Type :
- Report
- Accession number :
- edsarx.1404.1246
- Document Type :
- Working Paper
- Full Text :
- https://doi.org/10.1016/j.nima.2014.12.071