Back to Search
Start Over
Behavior of a trapezoid-based data acquisition system up to 100 kHz and beyond
- Source :
- Nuclear Instruments and Methods in Physics Research A 768 (2014) 192-196
- Publication Year :
- 2016
-
Abstract
- In this work, we investigated the ability of a high-purity germanium detector connected to a trapezoid-filter-based data acquisition system to reliably record signals in spite of high sample activities. By activating multiple Na$_{2}$CO$_{3}$ samples with different Na content, we were able to deduce efficiency, resolution and dead time of the system used as a function of the sample activity. Based on the results, we were able to find a setting which allows measurements of event rates up to 35~kHz per readout channel with an energy resolution of 0.3\% at the 2754 keV $^{24}$Na line.<br />Comment: 18 pages, 10 figures
- Subjects :
- Physics - Instrumentation and Detectors
Nuclear Experiment
Subjects
Details
- Database :
- arXiv
- Journal :
- Nuclear Instruments and Methods in Physics Research A 768 (2014) 192-196
- Publication Type :
- Report
- Accession number :
- edsarx.1611.09739
- Document Type :
- Working Paper
- Full Text :
- https://doi.org/10.1016/j.nima.2014.09.051