Back to Search
Start Over
Towards Picogram Detection of Superparamagnetic Iron-Oxide Particles Using a Gradiometric Receive Coil
- Publication Year :
- 2017
-
Abstract
- Superparamagnetic iron-oxide nanoparticles can be used in a variety of medical applications like vascular or targeted imaging. Magnetic particle imaging (MPI) is a promising tomographic imaging technique that allows visualizing the 3D nanoparticle distribution concentration in a non-invasive manner. The two main strengths of MPI are high temporal resolution and high sensitivity. While the first has been proven in the assessment of dynamic processes like cardiac imaging, it is unknown how far the detection limit of MPI can be lowered. Within this work, we will present a highly sensitive gradiometric receive-coil unit combined with a noise-matching network tailored for the measurement of mice. The setup is capable of detecting 5 ng of iron in vitro at 2.14 sec acquisition time. In terms of iron concentration we are able to detect 156 {\mu}g/L marking the lowest value that has been reported for an MPI scanner so far. In vivo MPI mouse images of a 512 ng bolus at 21.5 ms acquisition time allow for capturing the flow of an intravenously injected tracer through the heart of a mouse. Since it has been rather difficult to compare detection limits across MPI publications we propose guidelines improving the comparability of future MPI studies.<br />Comment: 15 Pages, 7 Figures, V2: Changed the initials of Author Kannan M Krishnan, added two citations, corrected typos
- Subjects :
- Physics - Medical Physics
Subjects
Details
- Database :
- arXiv
- Publication Type :
- Report
- Accession number :
- edsarx.1704.01789
- Document Type :
- Working Paper