Abstract
- We propose a straightforward sample-based technique to calibrate the axial detection in 3D single molecule localization microscopy (SMLM). Using microspheres coated with fluorescent molecules, the calibration curves of PSF-shaping- or intensity-based measurements can be obtained for any required depth range from a few hundreds of nanometers to several tens of microns. This experimental method takes into account the effect of the spherical aberration without requiring computational correction.<br />Comment: 8 pages, 2 figures. Submitted to Optics Letters on October 12th, 2017