The GlueX Start Counter Detector

Authors :
Pooser, Eric
Barbosa, Fernando
Boeglin, Werner
Hutton, Charles
Ito, Mark
Kamel, Mahmoud
Khetarpal, Puneet
Llodra, Anthony
Sandoval, Joseph
Taylor, Simon
Whitlatch, Timothy
Worthington, Stephanie
Yero, Carlos
Zihlmann, Benedikt
Source :
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2019
Publication Year :
2019

Abstract

The design, simulation, fabrication, calibration, and performance of the GlueX Start Counter detector is described. The Start Counter was designed to operate at integrated rates of up to 9 MHz with a timing resolution in the range of 500 to 825 ps (FWHM). The Start Counter provides excellent solid angle coverage, a high degree of segmentation for background rejection, and can be utilized in the level 1 trigger for the experiment. It consists of a cylindrical array of 30 thin scintillators with pointed ends that bend towards the beam line at the downstream end. Magnetic field insensitive silicon photomultiplier detectors were used as the light sensors.