Abstract
- The Jiangmen Underground Neutrino Observatory is proposed to determine neutrino mass hierarchy using a 20~ktonne liquid scintillator detector. Strict radio-purity requirements have been put forward for all the components of the detector. According to the MC simulation results, the radon dissolved in the water Cherenkov detector should be below 200~mBq/m$^3$. Radium, the progenitor of radon, should also be taken seriously into account. In order to measure the radium concentration in water, a radium measurement system, which consists of a radium extraction system, a radon emanation chamber and a radon concentration measurement system, has been developed. In this paper, the updated radon concentration in gas measurement system with a one-day-measurement sensitivity of $\sim$5~mBq/m$^3$, the detail of the development of the radium concentration in water measurement system with a sensitivity of $\sim$23~mBq/m$^3$ as well as the measurement results of Daya Bay water samples will be presented.<br />Comment: 17 pages, 10 figures and 3 tables
Details
- Database :
- arXiv
- Publication Type :
- Report
- Accession number :
- edsarx.1906.06895
- Document Type :
- Working Paper
- Full Text :
- https://doi.org/10.1016/j.nima.2020.164266