On low-energy tail distortions in the detector responsefunction of x-ray microcalorimeter spectrometers

Authors :
O'Neil, G. C.
Szypryt, P.
Takacs, E.
Tan, J. N.
Buechele, S. W.
Naing, A. S.
Joe, Y. I.
Swetz, D.
Schmidt, D. R.
Doriese, W. B.
Gard, J. D.
Reintsema, C. D.
Ullom, J. N.
Villarrubia, J. S.
Ralchenko, Yu.
Publication Year :
2020

Abstract

We use narrow spectral lines from the x-ray spectra of various highlycharged ions to measure low-energy tail-like deviations from a Gaussian responsefunction in a microcalorimter x-ray spectrometer with Au absorbers at energiesfrom 650 eV to 3320 eV. We review the literature on low energy tails in othermicrocalorimter x-ray spectrometers and present a model that explains all thereviewed tail fraction measurements. In this model a low energy tail arises fromthe combination of electron escape and energy trapping associated with Bi x-rayabsorbers.

Details

Database :
arXiv
Publication Type :
Report
Accession number :
edsarx.2005.05395
Document Type :
Working Paper
Full Text :
https://doi.org/10.1007/s10909-019-02270-y