Back to Search Start Over

Ultra-Low Background Germanium Assay at the Boulby Underground Laboratory

Authors :
Scovell, P. R.
Meehan, E.
Paling, S. M.
Thiesse, M.
Liu, X.
Ghag, C.
Ginsz, M.
Quirin, P.
Ralet, D.
Publication Year :
2023

Abstract

As we move to an era where next generation low-background particle physics experiments begin to be designed and constructed, the ability to assay materials with high sensitivity and at speed with a variety of techniques will be key. This paper describes the Mirion Technologies (Canberra) specialty ultra-low background detectors installed and commissioned at the Boulby Underground Laboratory between 2017 and 2021. The low background levels of the detectors combine with low background shielding and a radon-reduced dry nitrogen purge system to give sensitivity approaching the best in the world without the need for intricate shielding solutions. For an optimised sample geometry, run for 100 days, it would be possible to reach close to 10 $\mu$Bq/kg ($10^{-12}$ g/g) for background isotopes of interest in neutrinoless double-beta decay.

Details

Database :
arXiv
Publication Type :
Report
Accession number :
edsarx.2308.03444
Document Type :
Working Paper
Full Text :
https://doi.org/10.1088/1748-0221/19/01/P01017