Back to Search Start Over

Low Radon Concentration Measurement with a Radon-Dissolved Liquid Scintillator Pilot Detector

Authors :
Li, Qian-Yun
Li, Ming-Xuan
Li, Ren-Ming-Jie
Lin, Shin-Ted
Liu, Shu-Kui
Tang, Chang-Jian
Xing, Hao-Yang
Zhu, Jing-Jun
Yue, Qian
Yang, Li-Tao
Publication Year :
2023

Abstract

We construct a high sensitivity radon pilot detector using liquid scintillator dissolved radon for the CDEX rare-event searches program. The CDEX-10 project employs a germanium detector array immersed in a massive liquid nitrogen tank. However, radon emanated from the surface of the tank can contribute background. As a result, radon contamination in the liquid nitrogen tank must be regulated and monitored, which necessitates the use of a low-level Rn measuring device. The radon-dissolved liquid scintillator (LS) detectors utilizes cascade decay selections, yielding a unique signature. All background events occurring in the LS, with regard to the system and co-efficiency of Rn absorption, have been studied. Background activity is measured to be 59.8$\pm$18.4 $\mu$Bq per 300 mL LS. Meanwhile, it shows that radon concentration absorbed by LS will be saturated and is linearly related to radon concentration in nitrogen gas. Furthermore, the co-efficiency factor can be optimized by lowering temperature or raising pressure. Eventually the detection limit for 222^Rn in nitrogen gas is observed to be 9.6 mBq/m^3.<br />Comment: 8 pages, 18 figures

Details

Database :
arXiv
Publication Type :
Report
Accession number :
edsarx.2308.07160
Document Type :
Working Paper