Back to Search
Start Over
Sequence adaptive field-imperfection estimation (SAFE): retrospective estimation and correction of $B_1^+$ and $B_0$ inhomogeneities for enhanced MRF quantification
- Publication Year :
- 2023
-
Abstract
- $B_1^+$ and $B_0$ field-inhomogeneities can significantly reduce accuracy and robustness of MRF's quantitative parameter estimates. Additional $B_1^+$ and $B_0$ calibration scans can mitigate this but add scan time and cannot be applied retrospectively to previously collected data. Here, we proposed a calibration-free sequence-adaptive deep-learning framework, to estimate and correct for $B_1^+$ and $B_0$ effects of any MRF sequence. We demonstrate its capability on arbitrary MRF sequences at 3T, where no training data were previously obtained. Such approach can be applied to any previously-acquired and future MRF-scans. The flexibility in directly applying this framework to other quantitative sequences is also highlighted.<br />Comment: 12 pages, 5 figures, submitted to International Society for Magnetic Resonance in Medicine 31th Scientific Meeting, 2024
Details
- Database :
- arXiv
- Publication Type :
- Report
- Accession number :
- edsarx.2312.09488
- Document Type :
- Working Paper