A directional total variation minimization algorithm for isotropic resolution in digital breast tomosynthesis
- Publication Year :
- 2024
-
Abstract
- An optimization-based image reconstruction algorithm is developed for contrast enhanced digital breast tomosynthesis (DBT) using dual-energy scanning. The algorithm minimizes directional total variation (TV) with a data discrepancy and non-negativity constraints. Iodinated contrast agent (ICA) imaging is performed by reconstructing images from dual-energy DBT data followed by weighted subtraction. Physical DBT data is acquired with a Siemens Mammomat scanner of a structured breast phantom with ICA inserts. Results are shown for both directional TV minimization and filtered back-projection for reference. It is seen that directional TV is able to substantially reduce depth blur for the ICA objects.<br />Comment: Proceedings paper for accepted contribution to the 8th International Conference on Image Formation in X-Ray Computed Tomography (https://www.ct-meeting.org)