Abstract
- Rapid identification of microparticles in liquid is an important problem in environmental and biomedical applications such as for microplastic detection in water sources and physiological fluids. Existing spectro-scopic techniques are usually slow and not compatible with flow-through systems. Here we analyze single microparticles in the 14 - 20 micrometer range using a combination of two electronic sensors in the same microfluidic system: a microwave capacitive sensor and a resistive pulse sensor. Together, this integrated sen-sor system yields the effective electrical permittivity of the analyte particles. To simplify data analysis, 3D electrode arrangements were used instead of planar electrodes, so that the generated signal is unaffected by the height of the particle in the microfluidic channel. With this platform, we were able to distinguish between polystyrene (PS) and polyethylene (PE) microparticles. We showcase the sensitivity and speed of this tech-nique and discuss the implications for the future application of microwave cytometry technology in the en-vironmental and biomedical fields.<br />Comment: 7 pages, 4 figures