Effect of Fabrication Technology on the Accuracy of Surgical Guides for Dental-Implant Surgery

Authors :
Lucio Lo Russo
Laura Guida
Pierluigi Mariani
Vincenzo Ronsivalle
Crescenzio Gallo
Marco Cicciù
Luigi Laino
Source :
Bioengineering, Vol 10, Iss 7, p 875 (2023)
Publication Year :
2023
Publisher :
MDPI AG, 2023.

Abstract

Background: The accuracy of surgical guides is a relevant factor in both surgical safety and prosthetic implications. The impact of widespread fabrication technologies (milling and 3D printing) was investigated. Methods: Surgical guides manufactured by means of two specific milling and 3D-printing systems were digitized and compared in a 3D analysis with the digital file of the designed guides. The surface mean 3D distance (at the surface where the teeth and mucosa made contact) and the axial and linear deviations of the sleeves’ housings were measured by means of a metrological software program. Univariate and multivariate statistical analyses were used to investigate the effects of the fabrication technology, type of support, and arch type on the surgical guides’ accuracy. Results: The median deviations of the intaglio surface in contact with the mucosa were significantly (p < 0.001) lower for the milled surgical guides (0.05 mm) than for the 3D-printed guides (−0.07 mm), in comparison with the reference STL file. The generalized estimated equation models showed that the axial deviations of the sleeves’ housings (a median of 0.82 degrees for the milling, and 1.37 degrees for the 3D printing) were significantly affected by the fabrication technology (p = 0.011) (the milling exhibited better results), the type of support (p < 0.001), and the combined effect of the fabrication technology and the sleeve-to-crest angle (p = 0.003). The linear deviation (medians of 0.12 mm for the milling and 0.21 mm for the 3D printing) of their center points was significantly affected by the type of support (p = 0.001), with the milling performing slightly better than the 3D printing. Conclusions: The magnitude of the difference might account for a limited clinical significance.

Details

Language :
English
ISSN :
23065354
Volume :
10
Issue :
7
Database :
Directory of Open Access Journals
Journal :
Bioengineering
Publication Type :
Academic Journal
Accession number :
edsdoj.01d744610e409dad085cd5f56446b7
Document Type :
article
Full Text :
https://doi.org/10.3390/bioengineering10070875