Concordance between four PD-L1 immunohistochemical assays and 22C3 pharmDx assay in esophageal squamous cell carcinoma in a multicenter study

Authors :
Liyan Xue
Yuan Li
Lili Jiang
Chao Liu
Na Cheng
Changyuan Guo
Yan Jin
Ping Zhou
Xuemin Xue
Yue Wang
Weiya Wang
Yanhui Liu
Jianming Ying
Source :
Journal of the National Cancer Center, Vol 4, Iss 2, Pp 162-168 (2024)
Publication Year :
2024
Publisher :
Elsevier, 2024.

Abstract

Background: The prediction of response to immunotherapy mostly depends on the programmed death-ligand 1 (PD-L1) immunohistochemistry (IHC) status, and the 22C3 pharmDx assay has been approved in esophageal squamous cell carcinoma (ESCC). However, the widespread use of the 22C3 pharmDx assay is limited due to its availability. Thus, alternative PD-L1 assays are needed. We aimed to investigate the analytical and clinical diagnostic performances of four PD-L1 assays and to compare their concordances with the 22C3 pharmDx assay. Methods: The PD-L1 22C3 pharmDx assay was performed on the Dako Autostainer Link 48 platform, three testing assays (PD-L1 E1L3N XP antibody [Ab], PD-L1 BP6099 Ab and PD-L1 CST E1L3N Ab) on the Leica BOND-MAX/III platform, and one testing assay (PD-L1 MXR006 Ab) on the Roche VENTANA Benchmark Ultra platform. A total of 218 ESCC cases from four centers were included in this retrospective study. Professionals from each center stained and read the IHC slides independently and determined the combined positive score (CPS) and the tumor proportion score (TPS). Results: Regarding analytical performance, the four testing assays demonstrated good correlations with the 22C3 pharmDx assay when evaluated by the TPS or CPS (ρ > 0.8 for all four assays). Regarding diagnostic performance (CPS ≥ 10 was used as the cutoff), the four testing assays showed moderate concordances with the 22C3 pharmDx assay (kappa > 0.7 for all four assays). The overall percent agreements between each testing assay and the 22C3 pharmDx assay was at least 87.2 %. Conclusion: This study provides insight into the potential interchangeability of the four PD-L1 assays with the 22C3 pharmDx assay.

Details

Language :
English
ISSN :
26670054
Volume :
4
Issue :
2
Database :
Directory of Open Access Journals
Journal :
Journal of the National Cancer Center
Publication Type :
Academic Journal
Accession number :
edsdoj.05ddd10e3504c71a2ec3c4f4d547b9d
Document Type :
article
Full Text :
https://doi.org/10.1016/j.jncc.2023.11.003