Back to Search Start Over

Skeletonized internal thoracic artery harvesting: a low thermal damage electrosurgical device provides improved endothelial layer and tendency to better integrity of the vessel wall compared to conventional electrosurgery

Authors :
Alicja Zientara
Paul Komminoth
Burkhardt Seifert
Dragan Odavic
Omer Dzemali
Achim Häussler
Michele Genoni
Source :
Journal of Cardiothoracic Surgery, Vol 13, Iss 1, Pp 1-8 (2018)
Publication Year :
2018
Publisher :
BMC, 2018.

Abstract

Abstract Background Electrosurgery is fundamental to the precise, fast and bloodless preparation of internal thoracic artery grafts in cardiac surgery. The PEAK PlasmaBlade is a monopolar electrosurgical device that uses pulsed radiofrequency energy to generate a plasma-mediated discharge along an insulated electrode, creating a cutting edge while the blade stays near body temperature. The aim of this study is to compare the histological samples, cardiac computed-tomography of graft patency, and clinical outcomes of patients after off-pump coronary artery bypass grafting with preparation of the internal thoracic arteries by a conventional electrosurgical device and the PlasmaBlade. Methods In twenty subjects one internal thoracic artery was prepared with PlasmaBlade and the other artery with a conventional electrosurgical device. Histological samples were evaluated for three factors for potential graft failure: endothelial damage, integrity of the vessel wall and adventitial hemorrhage. Five samples per artery were evaluated by a novel scoring method based on the exposed circumference of the histological sample (“0”: 0%, “1”: 1–25%, “2”: 26–50%, “3”: 51–75%, “4”: ≥76% of the circumference). The Wilcoxon signed ranks test for mean scores within subjects was performed. Six-month-follow up by cardiac computed tomography for evaluation of graft patency was completed in 16 patients. Results Histological results demonstrated significantly less endothelial damage after PlasmaBlade (83% vs 60%, absolute: 75/90 vs. 53/89 samples with score “0–1”, p = 0.04). PlasmaBlade samples demonstrated a tendency to better wall integrity (72% vs. 54%, absolute: 64/89 vs. 47/87 samples with score “0–1”, p = 0.32). There were no differences in endothelial bleeding (PlasmaBlade 46% vs. electrosurgery 53%, absolute: 41/88 vs. 48/90 samples with score “0–1”, p = 0.63). Computed tomography confirmed non-inferiority of the PlasmaBlade to conventional electrosurgery with a patency rate of 94%. Conclusion Histologically, internal thoracic arteries harvested with PlasmaBlade demonstrate a more intact endothelial layer and a tendency to better wall integrity. Computed tomography of graft patency speaks for non-inferiority to conventional electrosurgery. PlasmaBlade may be preferable to conventional electrosurgery, if further follow-up confirms patency of internal thoracic arteries. Trial registration NCT03510026, registered 4th April 2018 (retrospectively registered).

Details

Language :
English
ISSN :
17498090
Volume :
13
Issue :
1
Database :
Directory of Open Access Journals
Journal :
Journal of Cardiothoracic Surgery
Publication Type :
Academic Journal
Accession number :
edsdoj.0b6c8043cd324247a6169dbff07d21c4
Document Type :
article
Full Text :
https://doi.org/10.1186/s13019-018-0797-3