Effect of scanning parameters on dose-response of radiochromic films irradiated with photon and electron beams

Authors :
Nashrulhaq Tagiling
Raizulnasuha Ab Rashid
Siti Nur Amirah Azhan
Norhayati Dollah
Moshi Geso
Wan Nordiana Rahman
Source :
Heliyon, Vol 4, Iss 10, Pp e00864- (2018)
Publication Year :
2018
Publisher :
Elsevier, 2018.

Abstract

Proper dosimetry settings are crucial in radiotherapy to ensure accurate radiation dose delivery. This work evaluated scanning parameters as affecting factors in reading the dose-response of EBT2 and EBT3 radiochromic films (RCFs) irradiated with clinical photon and electron beams. The RCFs were digitised using Epson® Expression® 10000XL flatbed scanner and image analyses of net optical density (netOD) were conducted using five scanning parameters i.e. film type, resolution, image bit depth, colour to grayscale transformation and image inversion. The results showed that increasing spatial resolution and deepening colour depth did not improve film sensitivity, while grayscale scanning caused sensitivity reduction below than that detected in the Red-channel. It is also evident that invert and colour negative film type selection negated netOD values, hence unsuitable for scanning RCFs. In conclusion, choosing appropriate scanning parameters are important to maintain preciseness and reproducibility in films dosimetry.

Details

Language :
English
ISSN :
24058440
Volume :
4
Issue :
10
Database :
Directory of Open Access Journals
Journal :
Heliyon
Publication Type :
Academic Journal
Accession number :
edsdoj.0cf8fbf5f4ea41348124a558f19fc4ff
Document Type :
article
Full Text :
https://doi.org/10.1016/j.heliyon.2018.e00864