Back to Search
Start Over
An Easily Overlooked Contamination of Syringes in Newborn Screening by Tandem Mass Spectrometry
- Source :
- Frontiers in Pediatrics, Vol 8 (2021)
- Publication Year :
- 2021
- Publisher :
- Frontiers Media S.A., 2021.
-
Abstract
- Background: Tandem mass spectrometry becomes a common and important test in newborn screening, but potential contamination of the equipment has largely been ignored.Methods: The source of contamination through Biosan quality control samples was examined prospectively, and further confirmed by retrospective analysis of patient samples.Results: We found that the source of contamination came from a syringe in the Biosan quality control samples. Furthermore, we found that a large number of indicators in the patient sample were interfered by syringe contamination in our center, and also in two other newborn screening centers, but the affected indicators were different in different screening centers.Conclusion: Syringe contamination will affect the detection of patient samples by tandem mass spectrometry and should be monitored carefully and immediately.
Details
- Language :
- English
- ISSN :
- 22962360
- Volume :
- 8
- Database :
- Directory of Open Access Journals
- Journal :
- Frontiers in Pediatrics
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.10c5ec1fc4b64f94a04004d51af4d3b8
- Document Type :
- article
- Full Text :
- https://doi.org/10.3389/fped.2020.596321