Back to Search Start Over

High-throughput Quant-iT PicoGreen assay using an automated liquid handling system

Authors :
Kay Anantanawat
Nicola Pitsch
Caroline Fromont
Caroline Janitz
Source :
BioTechniques, Vol 66, Iss 6, Pp 290-294 (2019)
Publication Year :
2019
Publisher :
Taylor & Francis Group, 2019.

Abstract

Workflows in NGS facilities require high-standard practices and high-throughput pipelines to process the large number of samples received in a timely manner. Downstream protocols such as NGS library preparation require accurate estimation of nucleic acid concentrations, which can be achieved using fluorescent dye-based nucleic acid measurement. Here, we report a protocol for preparing a 384-well Quant-iT PicoGreen assay. The protocol allows the concentrations of 184 DNA samples to be measured simultaneously in duplicate in only 1 h using an Eppendorf epMotion 5075 liquid handling system. The advantages of this high-throughput approach include a reduction in both reagents (10x less reagents compared to a standard protocol) and time (3 h for 384 samples compared with 3 days).

Details

Language :
English
ISSN :
19409818 and 07366205
Volume :
66
Issue :
6
Database :
Directory of Open Access Journals
Journal :
BioTechniques
Publication Type :
Academic Journal
Accession number :
edsdoj.1488878c8524f3c91f5ea742a82b11b
Document Type :
article
Full Text :
https://doi.org/10.2144/btn-2018-0172