Expiratory high-frequency percussive ventilation: a novel concept for improving gas exchange

Authors :
Ferenc Peták
Gergely H. Fodor
Álmos Schranc
Roberta Südy
Ádám L. Balogh
Barna Babik
André Dos Santos Rocha
Sam Bayat
Davide Bizzotto
Raffaele L. Dellacà
Walid Habre
Source :
Respiratory Research, Vol 23, Iss 1, Pp 1-11 (2022)
Publication Year :
2022
Publisher :
BMC, 2022.

Abstract

Abstract Background Although high-frequency percussive ventilation (HFPV) improves gas exchange, concerns remain about tissue overdistension caused by the oscillations and consequent lung damage. We compared a modified percussive ventilation modality created by superimposing high-frequency oscillations to the conventional ventilation waveform during expiration only (eHFPV) with conventional mechanical ventilation (CMV) and standard HFPV. Methods Hypoxia and hypercapnia were induced by decreasing the frequency of CMV in New Zealand White rabbits (n = 10). Following steady-state CMV periods, percussive modalities with oscillations randomly introduced to the entire breathing cycle (HFPV) or to the expiratory phase alone (eHFPV) with varying amplitudes (2 or 4 cmH2O) and frequencies were used (5 or 10 Hz). The arterial partial pressures of oxygen (PaO2) and carbon dioxide (PaCO2) were determined. Volumetric capnography was used to evaluate the ventilation dead space fraction, phase 2 slope, and minute elimination of CO2. Respiratory mechanics were characterized by forced oscillations. Results The use of eHFPV with 5 Hz superimposed oscillation frequency and an amplitude of 4 cmH2O enhanced gas exchange similar to those observed after HFPV. These improvements in PaO2 (47.3 ± 5.5 vs. 58.6 ± 7.2 mmHg) and PaCO2 (54.7 ± 2.3 vs. 50.1 ± 2.9 mmHg) were associated with lower ventilation dead space and capnogram phase 2 slope, as well as enhanced minute CO2 elimination without altering respiratory mechanics. Conclusions These findings demonstrated improved gas exchange using eHFPV as a novel mechanical ventilation modality that combines the benefits of conventional and small-amplitude high-frequency oscillatory ventilation, owing to improved longitudinal gas transport rather than increased lung surface area available for gas exchange.

Details

Language :
English
ISSN :
1465993X
Volume :
23
Issue :
1
Database :
Directory of Open Access Journals
Journal :
Respiratory Research
Publication Type :
Academic Journal
Accession number :
edsdoj.1d8ecf96a75d4746bfc54716acb6df49
Document Type :
article
Full Text :
https://doi.org/10.1186/s12931-022-02215-2