Back to Search
Start Over
Simplified swift and safe vascular closure device deployment without a local arteriogram: Single center experience in 2074 consecutive patients
- Source :
- Indian Heart Journal, Vol 68, Iss 4, Pp 529-538 (2016)
- Publication Year :
- 2016
- Publisher :
- Elsevier, 2016.
-
Abstract
- Objective: Vascular closure devices (VCDs), such as the Angio-Seal, a three-component hemostatic plug, have greatly facilitated the routine clinical practice in the catheterization laboratory. The manufacturer recommends a local angiogram before Angio-Seal deployment. However, from the outset, we employed a simplified routine of deploying this VCD, i.e. without use of local angiography. Methods: The Angio-Seal was employed without a preceding femoral arteriogram over 8 years in 2074 consecutive patients, 72% presenting with acute coronary syndromes and subjected to coronary angiography (n = 1032) or PCI n = 1042) via a transfemoral approach with use of heparin and dual antiplatelet therapy. Results: Deployment of the VCD was successful in 99.4%. Complete hemostasis was obtained in 98% of cases. In 14 patients, Angio-Seal deployment failed. Mean time for placement of Angio-Seal was
Details
- Language :
- English
- ISSN :
- 00194832
- Volume :
- 68
- Issue :
- 4
- Database :
- Directory of Open Access Journals
- Journal :
- Indian Heart Journal
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.23e2e5916c484c5fac83f1450f9fde1e
- Document Type :
- article
- Full Text :
- https://doi.org/10.1016/j.ihj.2015.11.036