Back to Search Start Over

Simplified swift and safe vascular closure device deployment without a local arteriogram: Single center experience in 2074 consecutive patients

Authors :
Antonis S. Manolis
Georgios Georgiopoulos
Dimitris Stalikas
Spyridon Koulouris
Source :
Indian Heart Journal, Vol 68, Iss 4, Pp 529-538 (2016)
Publication Year :
2016
Publisher :
Elsevier, 2016.

Abstract

Objective: Vascular closure devices (VCDs), such as the Angio-Seal, a three-component hemostatic plug, have greatly facilitated the routine clinical practice in the catheterization laboratory. The manufacturer recommends a local angiogram before Angio-Seal deployment. However, from the outset, we employed a simplified routine of deploying this VCD, i.e. without use of local angiography. Methods: The Angio-Seal was employed without a preceding femoral arteriogram over 8 years in 2074 consecutive patients, 72% presenting with acute coronary syndromes and subjected to coronary angiography (n = 1032) or PCI n = 1042) via a transfemoral approach with use of heparin and dual antiplatelet therapy. Results: Deployment of the VCD was successful in 99.4%. Complete hemostasis was obtained in 98% of cases. In 14 patients, Angio-Seal deployment failed. Mean time for placement of Angio-Seal was

Details

Language :
English
ISSN :
00194832
Volume :
68
Issue :
4
Database :
Directory of Open Access Journals
Journal :
Indian Heart Journal
Publication Type :
Academic Journal
Accession number :
edsdoj.23e2e5916c484c5fac83f1450f9fde1e
Document Type :
article
Full Text :
https://doi.org/10.1016/j.ihj.2015.11.036