A successful retrieval of stripped outer coating of J-tip diagnostic guidewire from the left popliteal artery during elective coronary angiography

Authors :
Damjanović Miodrag
Šalinger-Martinović Sonja
Đorđević-Radojković Danijela
Koraćević Goran
Miloradović Vladimir
Source :
Vojnosanitetski Pregled, Vol 71, Iss 10, Pp 969-971 (2014)
Publication Year :
2014
Publisher :
Military Health Department, Ministry of Defance, Serbia, 2014.

Abstract

Introduction. Entrapment and fracture of diagnostic or therapeutic devices within the coronary circulatory system are a rare, but increasing problem. Case report. A 70-yearold man was admitted in our clinic for coronary angiography before the planned aortic valve replacement. An arterial sheath was inserted in the right common femoral artery. After introducing a J-tip diagnostic coronary guidewire into the aorta and advancing a left Judkins diagnostic catheter over it, suddenly occured peeling off of the wire´s hydrophilic coating at the aortic arch level. Very soon, this outer coating of guidewire carried by the blood stream was entered into the left femoral artery, then into the left popliteal artery. This stripped part of guidewire was successfully caught and extracted out by using a goose-neck snare catheter. Conclusion. A sudden stripping of outer coating of a J-tip diagnostic hydrophilic coronary guidewire during coronary angiography is possible to manage quickly and successfully by the use of a simple cathether.

Details

Language :
English, Serbian
ISSN :
00428450
Volume :
71
Issue :
10
Database :
Directory of Open Access Journals
Journal :
Vojnosanitetski Pregled
Publication Type :
Academic Journal
Accession number :
edsdoj.262b53f0b44640dd8882a72c20ffd687
Document Type :
article
Full Text :
https://doi.org/10.2298/VSP1410969D