Infra-Red Imaging to Detect Respirator Leak in Healthcare Workers During Fit-Testing Clinic

Authors :
Darius Chapman
Campbell Strong
Kathryn D Tiver
Dhani Dharmaprani
Even Jenkins
Anand N Ganesan
Source :
IEEE Open Journal of Engineering in Medicine and Biology, Vol 5, Pp 198-204 (2024)
Publication Year :
2024
Publisher :
IEEE, 2024.

Abstract

Objective: This study addressed the problem of objectively detecting leaks in P2 respirators at point of use, an essential component for healthcare workers' protection. To achieve this, we explored the use of infra-red (IR) imaging combined with machine learning algorithms on the thermal gradient across the respirator during inhalation. Results: The study achieved high accuracy in predicting pass or fail outcomes of quantitative fit tests for flat-fold P2 FFRs. The IR imaging methods surpassed the limitations of self fit-checking. Conclusions: The integration of machine learning and IR imaging on the respirator itself demonstrates promise as a more reliable alternative for ensuring the proper fit of P2 respirators. This innovative approach opens new avenues for technology application in occupational hygiene and emphasizes the need for further validation across diverse respirator styles. Significance Statement: Our novel approach leveraging infra-red imaging and machine learning to detect P2 respirator leaks represents a critical advancement in occupational safety and healthcare workers' protection.

Details

Language :
English
ISSN :
26441276
Volume :
5
Database :
Directory of Open Access Journals
Journal :
IEEE Open Journal of Engineering in Medicine and Biology
Publication Type :
Academic Journal
Accession number :
edsdoj.280da26bdc040298cd2fd5da1abac87
Document Type :
article
Full Text :
https://doi.org/10.1109/OJEMB.2023.3330292