Abstract Objective Functional endoscopic sinus surgery is a commonly performed otolaryngologic procedure that often uses the microdebrider device for tissue removal. Given the ubiquitous nature of the instrument, we sought to better define the patterns of device failure using the postmarket surveillance openFDA database. Methods The openFDA database was queried for all microdebrider‐related adverse events from January 1, 2000 to November 1, 2020. Descriptive information on the nature of device failure and any associated patient injury was compiled. Reports not directly related to device failure were excluded from the analysis. Results A total of 641 events were included in the analysis. The most common device failure was overheating (n = 348, 54.3%), followed by material separation (n = 173, 27%), and inconsistent device activation (n = 52, 8.1%). Of the reported events, the vast majority did not result in patient harm (n = 579, 90.3%). On review of the remaining cases, only 24 events (3.7%) resulted in true harm to the patient, defined as a temporary or permanent injury or >30 min of additional anesthesia time. Of these cases, the need to reschedule surgical cases (n = 5, 0.8%), retained foreign body (n = 5, 0.8%), and thermal tissue injury (n = 3,0.5%) were the most common. Five patients suffered an injury due to surgeon error unrelated to device malfunction (n = 5, 0.8%). Conclusions Microdebrider device failures are extremely rare. When they do occur, less than 10% result in patient harm. In cases of patient harm related to microdebrider failure, preoperative testing of the device before use could prevent many of the reported malfunctions.

Details

Language :
English
ISSN :
25891081 and 20958811
Volume :
9
Issue :
4
Database :
Directory of Open Access Journals
Journal :
World Journal of Otorhinolaryngology-Head and Neck Surgery
Publication Type :
Academic Journal
Accession number :
edsdoj.2accab441ffa4314aa54ff236dec8c67
Document Type :
article
Full Text :
https://doi.org/10.1002/wjo2.89