Back to Search
Start Over
Repeatability of central corneal thickness measurement with the Pentacam HR system
- Source :
- Revista Brasileira de Oftalmologia, Vol 71, Iss 1, Pp 14-17 (2012)
- Publication Year :
- 2012
- Publisher :
- Sociedade Brasileira de Oftalmologia, 2012.
-
Abstract
- PURPOSE: To assess the repeatability of central corneal thickness measurement at the geometrical center (Central Corneal Thickness - CCT) given by the Pentacam High Resolution (HR) Comprehensive Eye Scanner (Oculus, Wetzlar, Germany) over time. METHODS: Prospective, single center, observational study. Two separate CCT measurements were taken by the Pentacam corneal tomography exam (CTm) 3 to 12 months apart, and compared. RESULTS: One hundred and sixteen eyes (n=116) of 62 health patients were included in this study. Average CCT in first and last visits was 541.6±37 µm and 543.6±36.9 µm respectively. Mean difference between both measurements was 9.2±6.4 µm, and there was no statistically significant difference in CCT measurement between visits, with good correlation between them (P = 0.057, r² = 0,9209). CONCLUSION: Pentacam (HR) CTm gives repeatable CCT measurements over time.
Details
- Language :
- English, Spanish; Castilian, Portuguese
- ISSN :
- 19828551 and 00347280
- Volume :
- 71
- Issue :
- 1
- Database :
- Directory of Open Access Journals
- Journal :
- Revista Brasileira de Oftalmologia
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.35dbab7cb93543ffbf8b50914291ee45
- Document Type :
- article
- Full Text :
- https://doi.org/10.1590/S0034-72802012000100003