Back to Search
Start Over
Technical Aspects of Coenzyme Q10 Analysis: Validation of a New HPLC-ED Method
- Source :
- Antioxidants, Vol 11, Iss 3, p 528 (2022)
- Publication Year :
- 2022
- Publisher :
- MDPI AG, 2022.
-
Abstract
- The biochemical measurement of the CoQ status in different tissues can be performed using HPLC with electrochemical detection (ED). Because the production of the electrochemical cells used with the Coulochem series detectors was discontinued, we aimed to standardize a new HPLC-ED method with new equipment. We report all technical aspects, troubleshooting and its performance in different biological samples, including plasma, skeletal muscle homogenates, urine and cultured skin fibroblasts. Analytical variables (intra- and inter-assay precision, linearity, analytical measurement range, limit of quantification, limit of detection and accuracy) were validated in calibrators and plasma samples and displayed adequate results. The comparison of the results of a new ERNDIM external quality control (EQC) scheme for the plasma CoQ determination between HPLC-ED (Lab 1) and LC-MS/MS (Lab 2) methods shows that the results of the latter were slightly higher in most cases, although a good consistency was generally observed. In conclusion, the new method reported here showed a good analytical performance. The global quality of the EQC scheme results among different participants can be improved with the contribution of more laboratories.
Details
- Language :
- English
- ISSN :
- 20763921
- Volume :
- 11
- Issue :
- 3
- Database :
- Directory of Open Access Journals
- Journal :
- Antioxidants
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.4535e86021640f2921d1a358eac4946
- Document Type :
- article
- Full Text :
- https://doi.org/10.3390/antiox11030528