Validation of an automated sample preparation module directly connected to LC-MS/MS (CLAM-LC-MS/MS system) and comparison with conventional immunoassays for quantitation of tacrolimus and cyclosporin A in a clinical setting

Authors :
Tsutomu Shimada
Daisuke Kawakami
Arimi Fujita
Rintaro Yamamoto
Satoshi Hara
Kiyoaki Ito
Ichiro Mizushima
Shinji Kitajima
Yasunori Iwata
Norihiko Sakai
Mitsuhiro Kawano
Takashi Wada
Yoshimichi Sai
Source :
Journal of Pharmaceutical Health Care and Sciences, Vol 10, Iss 1, Pp 1-7 (2024)
Publication Year :
2024
Publisher :
BMC, 2024.

Abstract

Abstract Background Therapeutic drug monitoring (TDM) systems generally use either liquid chromatography/tandem mass spectrometry (LC-MS/MS) or immunoassay, though both methodologies have disadvantages. In this study, we aimed to evaluate whether a CLAM-LC-MS/MS system, which consists of a sample preparation module directly connected to LC-MS/MS, could be used for clinical TDM work for immunosuppressive drugs in whole blood, which requires a hemolytic process. For this purpose, we prospectively validated this system for clinical measurement of tacrolimus and cyclosporin A in patients’ whole blood. The results were also compared with those of commercial immunoassays. Methods Whole blood from patients treated with tacrolimus or cyclosporin A at the Department of Nephrology and Departments of Rheumatology, Kanazawa University Hospital, from May 2018 to July 2019 was collected with informed consent, and drug concentrations were measured by CLAM-LC-MS/MS and by chemiluminescence immunoassay (CLIA) for tacrolimus and affinity column-mediated immunoassay (ACMIA) for cyclosporin A. Correlations between the CLAM-LC-MS/MS and immunoassay results were analyzed. Results Two hundred and twenty-four blood samples from 80 patients were used for tacrolimus measurement, and 76 samples from 21 patients were used for cyclosporin A. Intra- and inter-assay precision values of quality controls were less than 7%. There were significant correlations between CLAM-LC-MS/MS and the immunoassays for tacrolimus and cyclosporin A (Spearman rank correlation coefficients: 0.861, 0.941, P

Details

Language :
English
ISSN :
20550294
Volume :
10
Issue :
1
Database :
Directory of Open Access Journals
Journal :
Journal of Pharmaceutical Health Care and Sciences
Publication Type :
Academic Journal
Accession number :
edsdoj.4cbdbd40e3974201aab265833e9ec354
Document Type :
article
Full Text :
https://doi.org/10.1186/s40780-023-00318-6