Back to Search Start Over

Upgrade of Thomson Scattering Diagnostic on HL-2A

Authors :
Wenping Guo
Yuan Huang
Chunhua Liu
Zhen Feng
Zhipei Hou
Wenyan Zhai
Hisamichi Funaba
Ichihiro Yamada
Yonggao Li
Zhongbin Shi
Source :
Instruments, Vol 7, Iss 1, p 12 (2023)
Publication Year :
2023
Publisher :
MDPI AG, 2023.

Abstract

The Thomson scattering diagnostic of the HL-2A tokamak device was upgraded to improve its multi-point diagnostic capability, including new collection optics, fibers bundles, and data analysis code. The small old collection lens was replaced by a six-piece lens with a Cooke optical design. The aperture of its first standard sphere face is 310.125 mm, which successfully increases the amount of collected scattering light by about three times. The new collection optic module allows for up to twenty-six spatial points. A kind of Y-type fiber bundle has also been used to ensure that the fiber end-face matches the image of the laser beam exactly. Additionally, the new data analysis code can provide preview results in seconds. Finally, the multi-point Te diagnostic ability has been significantly improved.

Details

Language :
English
ISSN :
2410390X
Volume :
7
Issue :
1
Database :
Directory of Open Access Journals
Journal :
Instruments
Publication Type :
Academic Journal
Accession number :
edsdoj.4e1c6f3a4f0402e9c4605c0715db25f
Document Type :
article
Full Text :
https://doi.org/10.3390/instruments7010012