Back to Search
Start Over
Prevention of urinary tract infection using a silver alloy hydrogel-coated catheter in critically ill patients: A single-center prospective randomized controlled study
- Source :
- Journal of Intensive Medicine, Vol 4, Iss 1, Pp 118-124 (2024)
- Publication Year :
- 2024
- Publisher :
- Elsevier, 2024.
-
Abstract
- Background: A new type of silver alloy hydrogel-coated (SAH) catheter has been shown to prevent bacterial adhesion and colonization by generating a microcurrent, and to block the retrograde infection pathway. However, these have only been confirmed in ordinary patients. This study aims to evaluate the effectiveness of a SAH catheter for preventing urinary tract infections in critically ill patients. Methods: This was a prospective single-center, single-blind, randomized, controlled study. A total of 132 patients requiring indwelling catheterization in the intensive care unit (ICU) of the First Affiliated Hospital of the University of Science and Technology of China between October 2022 and February 2023 and who met the study inclusion/exclusion criteria were randomly divided into two groups. Patients in the SAH catheter group received a SAH catheter, while patients in the conventional catheter group received a conventional siliconized latex Foley catheter. The main outcome measure was the incidence of catheter-associated urinary tract infections (CAUTIs). Secondary outcome indicators included urine positivity for white blood cells and positive urine cultures on 3 days, 7 days, 10 days, and 14 days after catheterization, number of viable bacteria in the catheter biofilm on day 14, pathogenic characteristics of positive urine cultures, length of ICU stay, overall hospital stay, ICU mortality, and 28-day mortality. All the data were compared between the two groups. Results: A total of 68 patients in the conventional catheter group and 64 patients in the SAH catheter group were included in the study. On day 7 after catheter placement, the positivity rate for urinary white blood cells was significantly higher in the conventional catheter group than in the SAH catheter group (33.8% vs. 15.6%, P=0.016). On day 10, the rates of positive urine cultures (27.9% vs. 10.9%, P=0.014) and CAUTIs (22.1% vs. 7.8%, P=0.023) were significantly higher in the conventional catheter group than in the SAH catheter group. On day 14, the numbers of viable bacteria isolated from the catheter tip ([3.21±1.91]×106 colony-forming units [cfu]/mL vs. [7.44±2.22]×104 cfu/mL, P
Details
- Language :
- English
- ISSN :
- 2667100X
- Volume :
- 4
- Issue :
- 1
- Database :
- Directory of Open Access Journals
- Journal :
- Journal of Intensive Medicine
- Publication Type :
- Academic Journal
- Accession number :
- edsdoj.4eac5ab566f149889987b53b5324c91f
- Document Type :
- article
- Full Text :
- https://doi.org/10.1016/j.jointm.2023.06.003