Lowering the setting value of the esophageal endoscopic submucosal dissection device enabled tissue damage control in vitro porcine model

Authors :
Yukiko Yamaguchi
Masaya Uesato
Shohei Yonemoto
Tetsuro Maruyama
Ryuma Urahama
Hiroshi Suito
Takashi Kishimoto
Yuki Shiko
Yoshihito Ozawa
Yohei Kawasaki
Hisahiro Matsubara
Source :
Scientific Reports, Vol 12, Iss 1, Pp 1-7 (2022)
Publication Year :
2022
Publisher :
Nature Portfolio, 2022.

Abstract

Abstract One of the complications of esophageal endoscopic submucosal dissection (ESD) is postoperative stricture formation. Stenosis formation is associated with inflammation and fibrosis in the healing process. We hypothesized that the degree of thermal damage caused by the device is related to stricture formation. We aimed to reveal the relationship between thermal damage and setting value of the device. We energized a resected porcine esophagus using the ESD device (Flush Knife 1.5). We performed 10 energization points for 1 s, 3 s, and 5 s at four setting values of the device. We measured the amount of current flowing to the conducted points and the temperature and evaluated the effects of thermal damage pathologically. As results, the mean highest temperatures for 1 s were I (SWIFT Effect3 Wat20): 61.19 °C, II (SWIFT Effect3 Wat30): 77.28 °C, III (SWIFT Effect4 Wat20): 94.50 °C, and IV (SWIFT Effect4 Wat30): 94.29 °C. The mean heat denaturation areas were I: 0.84 mm2, II: 1.00 mm2, III: 1.91 mm2, and IV: 1.54 mm2. The mean highest temperature and mean heat denaturation area were significantly correlated (P

Subjects

Subjects :
Medicine
Science

Details

Language :
English
ISSN :
20452322
Volume :
12
Issue :
1
Database :
Directory of Open Access Journals
Journal :
Scientific Reports
Publication Type :
Academic Journal
Accession number :
edsdoj.59304a1c5e464eae3f16079a48f2f0
Document Type :
article
Full Text :
https://doi.org/10.1038/s41598-022-06533-9