Back to Search Start Over

Right ventricular lead sensing latency in pacemaker therapy

Authors :
Fani Zagkli
Nikoleta Kalovrenti
Panagiotis Patrinos
Panagiotis Chronopoulos
John Chiladakis
Source :
Journal of Arrhythmia, Vol 38, Iss 5, Pp 756-762 (2022)
Publication Year :
2022
Publisher :
Wiley, 2022.

Abstract

Abstract Background Pacemaker implantation involves intraoperative testing of ventricular sensing using a device called a pacing system analyzer (PSA). The value obtained is expected to correspond to those taken by the pacemaker after its implantation. This study determined the latency period for sensing intracardiac electrogram (EGM) by the right ventricular (RV) lead. Methods Patients without significant heart disease and underlying intrinsic atrioventricular (AV) conduction underwent Medtronic or Abbott dual‐chamber pacemaker implantation with the RV lead positioned on the mid‐septum. Real‐time sensing data were obtained through PSA and after pacemaker implantation to evaluate latency as the time interval Q‐VS between the onset of QRS on surface electrocardiogram and the sensed EGM by the RV lead. Results Of 157 patients, 105 had narrow QRS (

Details

Language :
English
ISSN :
18832148 and 18804276
Volume :
38
Issue :
5
Database :
Directory of Open Access Journals
Journal :
Journal of Arrhythmia
Publication Type :
Academic Journal
Accession number :
edsdoj.5b7ee1a525841c7bbf9be6731477270
Document Type :
article
Full Text :
https://doi.org/10.1002/joa3.12767